C. Continue to hold the ONLINE button for a few seconds. When the POWER light stops flashing,
release the ONLINE button, then momentarily press the PRINT button. The print cycle should begin,
but it may take a couple of minutes until the thermal print head warms up.
D. After the test print has printed successfully, turn the printer off.
Note: If the test print is unsuccessful, check the Troubleshooting section of the EP-1000
manual.
Network Connection
1. Locate the Ethernet converter and its power supply in the top foam of the EP-1000 shipping box.
2. Plug the Ethernet converter into the parallel port connector at the back of the EP-1000. Secure it with
the two clips.
3. Insert the power connector into the back of the Ethernet converter. Plug the Ethernet converter power
module into a 120 vac power source. Ensure that the Pwr indicator at the back of the converter is lit.
4. Connect a network cable to the back of the Ethernet converter as follows:
A. Network: Use the straight-through cable (blue cable provided with EP-1000) if the EP-1000 and the
Pegasys are part of a network.
B. Point-to-Point: Use the crossover cable (gray cable provided with the EP-1000) if the EP-1000 is to
connect directly to the Pegasys.
5. Verify that the orange Link indicator is lit.
EP-1000 Parallel Converter
Prints are sent to the EP-1000 through the network, but the Ethernet converter plugs into the printer’s
parallel interface port. Therefore, the parallel interface must be selected at the EP-1000:
1. Turn on power to the EP-1000. After it boots up there should be no red indicators on the top panel. If
there are, resolve the problem first. For example, be sure the door is closed and that a loaded paper
cassette is installed.
2. The two interface indicators are to the right of the EP-1000. If the uppermost light is lit, then the parallel
interface is selected and no further action is required. Go to the next section.
3. If the lower light is lit then the wrong interface is selected. Do the following to select the parallel
interface:
A. Momentarily press the ONLINE button. This will cause the selected interface light to flash, indicating
that the printer is offline.
B. Momentarily press the INTERFACE button. This will toggle the interface selection.
4. When the Parallel interface has been selected (top interface light is flashing) press the ONLINE button
to return the EP-1000 to the ready mode. The top interface light should now be lit continuously.
Ethernet Converter Setup
See also the Assigning an IP Address section of the Codonics Color Dye Sublimation Printers Ver. 7.0
Release Document.
1. Determine the IP address to be used by the EP-1000:
A. On a large network, obtain network address information (IP, subnet and gateway) from the System
Administrator.
B. On an isolated network, use any IP in the range 192.9.200.1 to 192.9.200.254 that is not currently in
use. For example, when installing two printers, valid IPs could be 192.9.200.100 and 192.9.200.101.
Find the Pegasys IP address by going to System Maintenance > System Utilities > Unix
Window and typing:
ypcat hosts
2. Note the Ethernet address of the converter which is found on the label of the converter. It is a 12-digit
hexadecimal number starting with 00-40-17.
3. Launch the converter configuration tool by going to a shell window and typing: